Publication detail
ANALYSIS OF THE MANUFACTURE PERTURBATIONS OF THE TOTAL HIP REPLACEMENT
KOUKAL, M.
English title
ANALYSIS OF THE MANUFACTURE PERTURBATIONS OF THE TOTAL HIP REPLACEMENT
Type
Paper in proceedings (conference paper)
Language
en
Original abstract
One type of total hip replacement function loss is acetabular cup loosening from the pelvic bone. This article examines manufacture deviations as one of the possible reasons for this kind of failure. Both dimensions and geometry manufacturing deviations of ceramic head and polyethylene cup were analyzed. We find that deviations in the variables analysed here affect considered values of contact pressure and frictional moment. Furthermore, contact pressure and frictional moment are quantities affecting replacement success and durability.
Keywords in English
Total hip replacement, FEM, roundness, contact pressure, frictional moment
Released
2017-05-15
Publisher
VUT v Brně
Location
Brno
ISBN
978-80-214-5497-2
ISSN
1805-8248
Book
Engineering Mechanics 2017
Volume
23
Pages from–to
506–509
Pages count
4
BIBTEX
@inproceedings{BUT156140,
author="Milan {Koukal}",
title="ANALYSIS OF THE MANUFACTURE PERTURBATIONS OF THE TOTAL HIP REPLACEMENT",
booktitle="Engineering Mechanics 2017",
year="2017",
journal="Engineering Mechanics",
volume="23",
number="ročník 23",
pages="506--509",
publisher="VUT v Brně",
address="Brno",
isbn="978-80-214-5497-2",
issn="1805-8248"
}