Detail publikace

ELECTROMAGNET FOR CAPTURE OF MAGNETIC BEADS FROM LARGE VOLUME INTO MICROFLUIDIC CHAMBRE CHIP

SVOBODOVÁ, Z. KUPČÍK, R. SMĚLÁ, D. ROUPEC, J. MACHÁČEK, O. NOVÁKOVÁ, L. BÍLKOVÁ, Z.

Anglický název

ELECTROMAGNET FOR CAPTURE OF MAGNETIC BEADS FROM LARGE VOLUME INTO MICROFLUIDIC CHAMBRE CHIP

Typ

Stať ve sborníku v databázi WoS či Scopus

Jazyk

en

Originální abstrakt

Immunomagnetic separation is method of choice besides filtration or centrifugation when a bioanalyte is to be specifically isolated form a larger volume. Here, home-made electromagnet with special, optimized construction, to fit to the microfluidic rhombic chip is presented. The experiments show how to capture 1 mg of superparamagnetic beads (SMB) from 20 mL syringe into 120 µL chamber in microfluidic chip. The captured beads can be easily transported to another microfluidic device for subsequent analysis. Various flow rates were tested and SMB were collected firstly from a phosphate buffer and subsequently from a cow's milk sample.

Klíčová slova anglicky

electromagnet, magnetic beads, microfluidic, chip

Vydáno

2018-10-16

Nakladatel

Institute of Analytical Chemistry of the CAS

Místo

Brno

ISBN

978-80-904959-5-1

Kniha

CECE 2018

Strany od–do

286–289

Počet stran

4

BIBTEX


@inproceedings{BUT153354,
  author="SVOBODOVÁ, Z. and KUPČÍK, R. and SMĚLÁ, D. and ROUPEC, J. and MACHÁČEK, O. and NOVÁKOVÁ, L. and BÍLKOVÁ, Z.",
  title="ELECTROMAGNET FOR CAPTURE OF MAGNETIC BEADS FROM LARGE VOLUME INTO MICROFLUIDIC CHAMBRE CHIP",
  booktitle="CECE 2018",
  year="2018",
  series="15th Internation Interdisciplinary Meeting on Bioanalysis",
  pages="286--289",
  publisher="Institute of Analytical Chemistry of the CAS",
  address="Brno",